Coding Field
Selection or range
Factory setting (bold)
Display Info
O3 (2)
Current output
simulation
lin = linear (1)
sim = simulation (2)
Tab
=
Tabelle (3)
The simulation is terminated by
selecting (1) or (3).
See O2 (1), O2 (3) for other
characteristics.
O321
Enter simulation
value
current value
0.00 ... 22.00 mA
The current value entered here is
output through the current output.
O3 (3)
Enter current
output table
(only for S version)
lin = linear (1)
sim = simulation (2)
Tab
=
Table (3)
Values may also be added or
changed at a later point in time.
The values entered are automatic-
ally sorted in ascending order by
current value.
See O2 (1), O2 (2) for other
characteristics.
O331
Selection of table
option
read
edit
O332
Enter number of
table value pairs
1
1...10
Enter the number of x and y
value pairs (measured value and
associated current value).
O333
Selection of table
value pair
1
1 ... No. Elem.
asign
O334 Enter x value
pH 0.00
pH –2.00 ... 16.00
0mV
–1500 ... 1500 mV
0.0 %
0.0 ... 100.0 %
x value = Measured value
determined by user.
O335 Enter y value
0.00 mA
0.00 ... 20.00 mA
y value = current value
determined by user to be
associated with O234.
O336
Enter whether or
not the table
status is okay
yes
no
Return to O3.
If status = “no”, set table
correctly (all previous settings
are kept) or back to
measurement mode
(table will be deleted).
Factory settings are printed in bold face;
base version does not include functions in
italic
.
sim O2
Sel.Type
10.20
O321
mA
Simulat.
O2
Sel.Type
table
O331
Sel.Table
read
O332
No.Elem.
1
O333
Sel.Elem.
1
O334
Meas.val.
0.00
pH
O335
mA value
0.00
mA
O336
Status ok
yes
Instrument configuration Liquisys M CPM 223 / 253
32 Endress+Hauser