5.10 Calibration
This function group is used to calibrate the
transmitter. Two types of calibration are
possible:
Calibration by means of measurement in
two calibration solutions with a known pH
value.
Calibration by entering daa for slope and
zero point
For redox measurement by entering the
mV value or two different % values.
Note:
If the calibration procedure is
aborted by pressing the PLUS and
MINUS keys at the same time
(return to C19, C25 or C36) or if
the calibration is faulty, then the
previous calibration data is
reinstated. A calibration error is
indicated by the ERR message
and flashing of the sensor symbol
on the display.
Repeat the calibration!
The instrument is automatically
switched to hold during calibration
(factory setting).
An offset entered by the user is
automatically cancelled when the
calibration is accepted.
If the slope or zero point is outside
the ranges defined in C16 and
C17, error 32 is activated for slope
or error 33 for zero. The electrode
should be then checked and
replaced if necessary.
Coding Field
Selection or range
Factory setting
Display Info
I2 Tag description Only for communication.
I2
@@@@@@@@
Tag
PM253E5B.CHP
Liquisys M CPM 223 / 253 Instrument configuration
Endress+Hauser 53