Dropout delay
entry
0...2000 s
0s
R225
Entry of alarm threshold
(as an absolute value)
150.0 °C
-20.0...+150.0 °C
R226
Entry of derivative action
time Tv
(0.0 = no D component)
0.0...999.9 min
0.0 min
R235
Selection of control
characteristic
inv = inverted;
dir = direct;
R236
Selection
freq = pulse frequency
Ien = pulse length
R237
Entry of min. ON time
t
0.1...5.0 s
on
0.3 s
R2310
Entry of max.
pulse frequency
60...180 1/min
120 1/min
R239
Entry of
pulse interval
0.5...999.9 s
10.0 s
R238
Dropout delay
entry
0...2000 s
0s
R215
Entry of alarm threshold
(as an absolute value)
; -2.00...16.00 pH
; -1500...1500 mV
; 0...100.0 %
16.00 pH
1500 mV
100.0 %
R216
Entry of derivative action
time Tv
(0.0 = no D component)
0.0...999.9 min
0.0 min
R265
Entry of min. ON time
t
0.1...5.0 s
on
0.3 s
R269
Entry of max.
pulse frequency
60...180 1/min
120 1/min
R268
Selection
freq = pulse frequency
len = pulse length
R266
Entry of
pulse interval
0.5...999.9 s
10.0 s
R267
Entry of post-rinse time
0...999 s
20 s
R255
Number of repeat cycles
0...5
0
R256
Set interval between
two cleaning cycles
(pause time)
1...7200 min
360 min
R257
Number of cleaning
cycles without
cleaning agent
0...9
0
R259
Entry of minimum
pause time
1...R357 min
120 min
R258
Order number
is displayed
Reference voltage
is displayed
S7
S11
Serial number
is displayed
Select
AC frequency
S8
S12
Perform instrument test
; displayno
S10
Reset intrument
(restore default values)
Sens = sensor data;
Facty = factory settings
no;
S9
P7
P8
P10
P9
P12
P11
Set lower
alarm threshold
pH -2 ... 16pH -2.00
Set upper
alarm threshold
pH -2 ... 16pH 16.00
Select process
monitoring
AC; CC; AC CC
AC!; CC!; ACCC!
Off;
Set max. perm. period
for lower limit exceeded
0 ... 2000 min60 min
Set max. perm. period
for upper limit exceeded
0 ... 2000 min120 min
Set monitoring
value
pH -2 ... 16pH 1.00
Enter basic load
0 40%
0%
Enter process type
Inlne
Batch
Enter process type
Inlne
Batch
R2311 R2312
R2610