Limit monitoring
This function monitors the measuring value for exceeding monitoring limits and triggers an
alarm.
Coding Field
Selection or range
Factory setting (bold)
Display Info
P
Function group
CHECK
Settings for measuring electrode
and process monitoring
P1
Switch SCS alarm
for measuring
electrode on or off
Off
On
The glass electrode is monitored
for glass breakage (Error no.
E071.) Response time approx. 30s.
SCS monitoring is inactive during
calibration.
P2
Switch SCS alarm
for reference
electrode on or off
Off
On
The reference electrode is
monitored for soiling and
blocking. (Error no. E152.)
Response time approx. 60s.
Only at A2 = sym.
P3
Enter SCS alarm
threshold for
reference
electrode
50.0 k
0.5 ... 50.0 k
The measuring value contains
also the impedance of the
medium.
The impedance of the reference
electrode increases with the
degree of soiling.
P4
Leakage current
display for IsFET
0.0 ... 9.9
µ
A
Only if A4 = IsFET and
P1 = on.
Display only.
Leakage currents > 0.4
µ
A
indicate a damage of the IsFET
sensor.
P5
Select process
monitoring
Off
Low
High
Lo Hi = low + high
Lo!
Hi!
LoHi!
Alarm signalling optionally with or
without simultaneous controller
switch-off.
xxxx = without controller
switch-off
xxxx! = with controller switch-off
P6 Enter error delay
0 s (min)
0 ... 2000 s (min)
Depending on your selection in
F2, you can enter the error delay
in min or s. Only after this does a
high or low limit violation cause
an alarm as per field P7 / P8.
P7
Enter lower alarm
threshold
-2.00 pH
-2.00 ... 16.00 pH
Omitted when P5 = Off.
Factory settings are printed in bold face;
base version does not include functions in
italic
.
P
CHECK
Off
P1
SCS Glass
Off
P2
SCS Ref
P3
k@
SCSThresh
50.0
0.0
P4
Leak Curr
P5
A.Thresh
Off
P6
s
Err.Delay
0
P7
pH
LowAlarm
-2.00
Instrument configuration Liquisys M CPM 223 / 253